Stanford SR630 Thermocouple Monitor

16 channel thermocouple monitor (Distributed only in Germany.)

Specifications in Brief

Thermocouple
Channels16
Thermocouple typesB, E, J, K, R, S, T
Display units°C, °F and K
Display resolution0.1 °C
Temperature displaysActual, Nominal, or Offset
Accuracy0.5 °C for J, K, E and T. 1.0 °C for R, S and B. Errors are for the SR630 only. Standard errors for thermocouple wire are 2 to 5 times the error due to the SR630.
Inputs
Channels16
Input typeIndependent, floating and differential
Input resistance10 MΩ between + and - terminals, >1 GΩ to ground
Input capacitance0.001 µF
Input bias current<100 pA
Input protection250 Vrms
Open circuit check250 µA
Common mode±200 VDC
Voltmeter
Full-scale display(±9.999, ±99.99 or ±999.9) mVDC, (±9.999 or ±99.99) VDC
Range selectAutomatic
Resolution±1 of least significant displayed digit
Offset±2 of least significant displayed digit
Gain accuracy0.05 %
Conversion rate10/s for 50 Hz line, 12/s for 60 Hz
Line rejection>100:1
Scanning and Data Logging
ScanningSelected channels will be scanned. Dwell time between scans is set from 10 s to 9999 s.
AlarmTemperature or voltage limit for each channel
Scan enableAll channels may be scanned or skipped
Printer outputVoltages, Temperatures, Time and Date as a list or in a graphical format
Data memoryLast 2000 measurements in battery backed-up memory
General
Analog outputsFour analog voltage outputs proportional to temperature of channels 1, 2, 3 and 4
Relay outputSwitching 10 A
Store and recallNine locations for instrument set-up
InterfacesRS-232, GPIB, and Centronics ports (standard). All instrument functions may be set and read via RS-232 or GPIB.
Power10 W, 100/120/220/240 VAC, 50/60 Hz
Dimensions8.5" × 3.5" × 13" (WHL)
Weight9 lbs.